Test 2-3: Low-Level Outputs Test - Agilent Technologies 75000 Series Service Manual

Summing amplifier/dac module
Hide thumbs Also See for 75000 Series:
Table of Contents

Advertisement

Test 2-3: Low-level Outputs Test

Description
Equipment Setup
Test Procedure
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The purpose of this test is to verify that the amplifier meets its accuracy
specifications for low-level (differential) outputs.
Set Source to: DCV
Set DMM to: DCV, 10 V range
Figure 2-3. Low-level Outputs Test Setup
1. Reset the amplifier:
* RST;* CLS
Repeat steps 2-5 for each entry listed in Table 2-1:
2. Connect the DMM to the amplifier output specified in Table 2-1
(see Figure 2-3).
3. Connect the Source to the amplifier input specified in Table 2-1
(see Figure 2-3).
4. Set the Source to the input voltage specified in Table 2-1.
5. Measure the amplifier output with the DMM. Record the results
in Table 2-4.
Reset amplifier and clear
status registers
Verification Tests 2-13