Operating Instructions Biological Safety Cabinet | MSC-Advantage
3
Installation
3.3

Correct location

Choose a draft-free location where the biological safety cabinet
does not interfere with the plant traffic.
Fig. 1: This figure shows preferred locations for biological safety
cabinets and unsuitable locations that are not in accordance
with the safety requirements.
Unsuitable locations: The locations [1], [2], and [3] are not
suitable because they are exposed to drafts from windows
and doors.
Location [5] is unsuitable because it is within range of plant
traffic and within the exhaust air range of a ventilation system
[4].
Preferred locations: The locations [6], [7], and [8] are correct
because they are in a draft-free section of the room and not ex-
posed to plant traffic.
A counterweight at the device backpanel moves synchronously
with the vertical movement of the front window. To prevent the
counterweight from jamming, the device backpanel should be
as close to the wall as possible.
3.4

Installation in series

When several devices are to be installed in series, please ob-
serve the following:
Make sure that vibrations cannot be transferred
between adjacent devices.
Exterior surfaces of the cabinets must always be ac-
cessible for cleaning and disinfection.
14
1
2
3
8
7
6
Device arrangement in the
operating room
4
5
Fig. 1