6
Handling and control
Status indicators, Fig. 12:
[8] The air system is operated at reduced capacity (the yellow LED
[9] Airflow is not steady (the red LED is illuminated),
[10]Airflow is steady (the green LED is illuminated),
[11] Front window is not in work position (the red LED is illuminated),
[12]Front window is in work position (the green LED is illuminated).
SmartFlow dispaly, Fig. 12:
The SmartFlow Indicator displays the compensation capability of the exhaust con-
trol, with window in working position and blower on. The LEDs that illuminate in
pairs indicate the following:
[18] green + [17] green
[16] yellow + [15] yellow
[[14] red + [13] red
NOTE – Compensation capability!
If the display is steady yellow-yellow, there is
a possibility that the filters of the device are
fully loaded. In this case, the technical service
should be informed.
6.1.2

Display during calibration routine

NOTE – Calibration routine!
The calibration routine is started only upon the
initial operation.
The routine run takes approximately 30 minutes. During this time, the diplay al-
ternatingly shows cal and the remaining time. If the calibration routine cannot
be started due to failures, the display shows cal permanently.
6.1.3

Display during UV disinfection

If a UV lamp is installed, the diplay alternatingly shows dIS and the remaining
disinfection time.
Operating Instructions Biological Safety Cabinet | MSC-Advantage
is illuminated),
sufficient compensation capability
compensation capability is depleted
window outside working position or air velocity
outside alarm range
31