7

Operation

7.1

Hygiene preparations for the sample chamber

The sample chamber surfaces and the accessories required for the work process
must be disinfected and cleaned in accordance with the hygiene guidelines set
forth for the application.
7.2

Loading the sample chamber

Installing auxiliaries:
1. Move the front window to the maximum opening position and switch the blo-
wers on.
2. Install auxiliaries within the workplate working range.
3. Move the front window to the work position (the green status indicator F
W
WINDOW IN
ORK POSITION
zed (the green status indicator A
CAUTION – Operational safety!
The personal and material protection is ensu-
red only if the airflow system of the device is
working properly.
If the alarm system issues failure messages
when the front window is in the work position,
stop all applications that may release harmful
aerosols!
4. Load the work surface with samples.
5. For work intermissions or for extended experimental phases without manual in-
tervention, switch the device to standby mode.
7.3

Response to failure messages

Failure messages are displayed in form of text/number combinations with a code
between ER 3 to ER 7 (see Section 6.1.4). If one of these codes is displayed,
contact the Technical Service of Thermo Fisher Scientific immediately.
To isolate the cause of the failure, the operating personnel must perform only
the following tests and measures:
Check to see if the exhaust air opening on top of the cabinet is blocked.
Ensure that the on-site exhaust air system is activated.
Close doors and windows in the laboratory to prevent drafts.
Switch off devices in the vicinity of the biological safety cabinet that cause air
turbulence or emit excessive heat.
Open flames in the sample chamber may impair airflow conditions.
is illuminated) and wait until the airflow has stabili-
is illuminated).
IRFLOW STEADY
Operating Instructions Biological Safety Cabinet | MSC-Advantage
RONT
35