Operating Instructions Biological Safety Cabinet | MSC-Advantage
9
Cleaning and decontamination
9.7

Cleaning the floorpan

Clean the floorpan using a solution of tepid water and commer-
cially available dishwashing agent.
1. Remove the workplates from the sample chamber.
2. Remove dirt residues and deposits thoroughly.
3. Wipe the floorpan clean using a clean cloth and plenty of
clear water.
4. Remove the cleaning liquid from the floorpan and wipe the
floorpan surfaces thoroughly clean.
5. Reinstall the workplates.
9.8

Cleaning the protective grid

Fig. 16: The multi-segment protective grid is retained in the
installed position between support surface and airduct wall by
its own tension.
1. To remove a grid segment [2], push the retaining tab [1]
down until the grid segment can be removed from below
the counterholder [3] of the wall.
2. To install the grid segment [2], first place it behind the fixed
points [4], then push the retaining tab [1] down and against
the counterholder [3] of the wall so that the counterholder
engages in the tab.
42
NOTE – Material residues!
After cleaning, make sure that all cleaning
materials have been removed completely from
the floorpan.
NOTE – Inlet air protection!
Do not operate the device without inlet air
protection. Prior to any start-up of the device,
make sure that the inlet air protection is in-
stalled!
Fig. 16
Protective grid segment