Operating Instructions Biological Safety Cabinet | MSC-Advantage
4
Description of the device
4.6

Device interfaces

Fig. 7: The standard equipment includes:
outlets for internal/external power supply,
bushings on both sides for cables, media valves, and hoses,
a communication port,
a connection to external alarm systems.
Power supply connection: The connection to the power supply system is es-
tablished via a cable with grounding plug [4] at the upper side of the device.
Device-integral power supply: The backpanel contains outlets [10] for the power
supply (overall maximal current: 5 A) of internal accessories.
2 outlets (model MSC 0.9, 1.2)
4 outlets (model MSC 1.5, 1.8)
Optionally, an outlet can be replaced with a disinfection adapter [9] for the con-
nection of a mobile UV device.
Two fuse holders for 5 A miniature fuses at the top of the front cover protect the
device-integral power supply: [1] for (L), [2] for (N).
Communication port: The top of the device side contains an RS232 connec-
tion [3] for a PC.
Bushings: The standard fittings are 3 bushings per side panel.
SmartPort [6] is provided to lead through cables or hoses for accessories required
in the utility chamber. To this end, the rubber grommet [5] is punched out in the
exact opening size to avoid possible contamination.
Media valves [8] are installed solely in the bushings [7].
Media valves suitable for installing in bushings can be supplied as an option.
Upon delivery of the device, the bushings are sealed.
22
Fig. 7
Supply interfaces