4
Description of the device
4.5

Sample chamber access

Fig. 6: The manually movable front window [1] made of multi-
layer security glass seals the safety sample chamber front side
up.
The sample chamber is accessible through various positions
of the front window.
For operation:
Work position with opening height A for access to the sample
chamber during the work process,
maximal opening height B for loading the sample chamber,
position C (closed state): closed front window and reduced
blower speed.
NOTE – Status indicator!
The two positions A and C are determined by
switches and output as status indicators on the
display (see Section 6.1.1).
For cleaning / maintenance:
SmartClean window cleaning position with opening height D
for cleaning and disinfecting the upper section of the front win-
dow. For this purpose, the front window can be lowered
beyond closed position C so that a sufficiently high gap
exists between the sample chamber ceiling and the front win-
dow upper edge.
This gap can be also used for replacing the sample chamber
illumination lamps.
CAUTION – Front window jamming!
Do not use force to move the front window.
If the window is jammed or sluggish, contact
Technical Service immediately.
Do not attempt to repair this problem!
Operating Instructions Biological Safety Cabinet | MSC-Advantage