Change from Remote to Local Operation

The Product uses three modes of operation:
Local
Remote
Remote with Lockout
Local mode is the default mode. Commands can be entered with the buttons on the
Product or with a computer. In Remote mode, the buttons are disabled and commands can
only be entered with a computer. Select [GO TO LOCAL] from the menu on the
Product display to restore button operation. In Remote with Lockout mode, the buttons
cannot be used.
To switch modes:
1. To start Remote mode, enter the serial command "REMOTE" at the computer
terminal.
2. To start Remote with Lockout mode, enter "REMOTE LOCKOUT" in either order.
3. To go back to Local Operation mode, enter "LOCAL" at the terminal. This
command also turns off LOCKOUT if it is on. For more information on commands,
see the "Remote Commands" section.

Command Use

The Product can be controlled by commands and queries. All commands can be entered
with upper or lower case. See the Remote Commands section for all available commands.
The commands are divided into these categories:
Calibrator Commands
Only the Product uses these commands.
For example:
VAL?
This command requests the values shown on the Product display.
Common Commands
Standard commands used by most devices. These commands always start with an
"*".
For example:
*IDN?
This command tells the Product to show its identification.
Query Commands
Commands that request information and always end with a "?".
For example:
FUNC?
This command shows the current modes of the Product displays.
Portable Pressure Calibrator
Remote Operation
31