Calibrator Commands - Fluke 3130 User Manual

Portable pressure calibrator
Hide thumbs Also See for 3130:
Table of Contents

Advertisement

Portable Pressure Calibrator
Remote Operation

Calibrator Commands

DAMP
Turns on or off the dampening function.
For example:
If you send DAMP ON, this will turn the dampening function on.
DAMP?
Shows the current state of the dampening function.
For example:
If you send DAMP?, this will show ON if the dampening function is on.
DISPLAY
Turns on or off the indicated display.
For example:
If you send DISPLAY LOWER, ON, this will turn on the lower display.
DISPLAY?
Shows the current state of the each of the displays.
For example:
If you send DISPLAY?, it will show ON, ON, ON if the all the displays are on.
FAULT?
Shows the error code number of an error that has occurred. The command can be entered
if the previous command did not do what it was meant to do.
For example, if a value for current output is entered that is larger than the supported
range (0-24 mA), FAULT? shows:
103 which is the code number for an entry over range.
Refer to Table 6 for more information about error code numbers.
ERROR _LOOP
Turns on or off loop power in percent error mode.
For example:
To set loop power on, send ERROR_LOOP ON.
37