Portable Pressure Calibrator
Remote Operation
HART_ON
Turns on the Hart resistor.
HART_OFF
Turns off the Hart resistor.
HART?
Shows the state of the Hart resistor.
For example:
If the Hart resistor was on, HART? shows ON.
HI_ERR
Sets the 100 % point for the percent error mode calculation in the current engineering
units.
For example:
To set the 100 % point to 100 psi, send HI_ERR 100.
HI_ERR?
Shows the 100 % point for the percent error mode calculation.
For example:
If the 100 % point is set to 100 psi, HI_ERR? shows 1.000000E+02, PSI .
IO_STATE
Sets the input/output/simulate state of the mA function of the Product. This does not put
the Product into mA if it is not in it already.
For example:
If the Product is in mA simulate modem sending IO_STATE MEASURE would put it
into measure mode.
IO_STATE?
Shows the input/output/simulate state of the mA function of the Product.
For example:
If the Product was in mA simulate mode, IO_STATE? shows SIM.
39