Portable Pressure Calibrator
Remote Operation
OUT
This command also switches the Product into mA output mode. A number and a unit
must be entered after the command.
For example:
OUT 5 MA sets the current output at 5 mA.
OUT?
Shows the output of the Product.
Using the above example, OUT? shows: 5.000000E-03, A.
PRES_UNIT
Used to set the pressure unit for the indicated display.
For example:
To set the pressure unit to psi on the lower display, send PRES_UNIT LOWER, PSI.
PRES_UNIT?
Shows the pressure unit used when measuring pressure for each of the three displays.
PUMP_LIMIT
Sets the approximate pressure in psi at which the pump will turn off.
For example:
PUMP_LIMIT 50 sets the approximate value that the pump will shut off at to 50 psi.
PUMP_LIMIT?
Shows the pump limit. Using the above example, PUMP_LIMIT? Shows:
50.000
REMOTE
Puts the Product into remote mode. From the remote mode the user can still use the
keypad to get back to local mode unless the command LOCKOUT was entered before
REMOTE. Then the keypad is completely locked out, and the user has to send the
LOCAL command to go to local operation.