%-Error Function

The Product has a special function which can calculate the error in the pressure value
from the mA value as a percentage of the 4 ... 20 mA loop span. The %-Error mode uses
all three screens and has a special menu structure. It shows pressure, mA and %-Error
simultaneously. See Figure 28.
3130
EXTERNAL
PRESSURE MODULE
F1
HOME
ZERO
LOOP
V mA
COM
SWITCH TEST
30 V MAX
Example:
If a pressure transmitter under test has a full-scale range of 2 bar and gives a 4 ... 20 mA
output signal, you can program in a 0 ... 2 bar pressure span into the Product. The Product
will then calculate and show the deviation or %-Error value from the 4 ... 20 mA output.
This makes manual calculations unnecessary.
To use the "%-ERROR" function:
1. Once the Product is turned on, push  to start the "MORE" menu option. Then Push
2. Push  to select the "CONFIG" option.
26
Fully Closed Close silver vent knob
PORTABLE PRESSURE CALIBRATOR
AIR SUPPLY
FILTER
330 PSI MAX
(2.3 MPa)
CLOSE VALVE BEFORE
PRESSURE
OPERATING PUMP
SENSOR
F2
F3
SUPPLY
METERING VALVE
PRESSURE
VENT
VENT
FINE
CAUTION
ADJUSTMENT
PUMP
FINE
VENT LINE BEFORE MAKING SELECTION
ADJUSTMENT
24 V DC
CHARGE
16 V DC
5.6 A
Figure 28. Connections with a Pressure Transmitter with % Error Function
 to start the "%-ERROR" option.
OUTPUT
-12 PSI TO 300 PSI
(-8 kPa TO 2 MPa)
CLOSED
ISOLATION
VALVE
VACUUM
PUMP
FLUKE CORPORATION
SERIAL NO.
EVERETT, WA USA
www.flukecal.com
Use low volume
hose if possible
+
hbb027.eps