Change Between Remote and Local Operation

There are three modes of operation of the Product: Local,
Remote, and Remote with Lockout. Local mode is the
default mode. Commands may be entered with the
keypad on the Product or with a computer. In Remote
mode the keypad is disabled and commands may only be
entered with a computer. If [GO TO LOCAL] is chosen
from the display menu, the Product restores keypad
operation. In Remote with Lockout, the keypad cannot be
used.
To switch modes:
1.
To enable Remote mode, enter the serial command
REMOTE at the computer terminal.
2.
To enable Remote with Lockout, enter "REMOTE
LOCKOUT" in either order.
3.
To switch back to local operation enter LOCAL at the
terminal. This command also turns off LOCKOUT if it
is on. For more information on commands, refer to
the "Remote Commands" section.
42

Command Use and Types

Refer to the "Remote Commands" section for all available
commands. The Product may be controlled by commands
and queries. All commands may be entered with upper or
lower case. The commands are divided into the
categories listed below:
Calibrator Commands
Only the Product uses these commands. For example:
VAL?
asks for the values displayed on the Product display.
Common Commands
Standard commands used by most devices. These
commands always begin with an "*".
For example:
*IDN?
tells the Product to return its identification.