Fluke 3130 User Manual page 50

Portable pressure calibrator
Hide thumbs Also See for 3130:
Table of Contents

Advertisement

3130
Users Manual
SIM
Sets the output for current simulation. This command also switches the Product into mA
simulation mode. A number and a unit must be entered after the command.
For example:
SIM 5 MA sets the current simulation at 5 mA
SIM?
Shows the output of the current simulation. With the example above, the output would be:
5.000000E-03, A
ST_START
Starts a switch test.
ST_CLOSE?
Shows the pressure that the switch closed at in the current pressure units.
ST_OPEN?
Shows the pressure that the switch opened at in the current pressure units.
ST_DEAD?
Shows deadband of the switch in the current pressure units.
VAL?
Shows the value of measurement on a maximum of three displays. For example, if the
upper display shows pressure at 0.00 bar on P1, the middle display shows pressure at
0 psi on P1, and the lower display shows 0 mA, then VAL? will return:
0.000000E+00,BAR,0.000000E+00,PSI,0.000000E+00,A
ZERO_MEAS
Zeroes the attached pressure module. Enter the zeroing value in PSI after the command
when zeroing an absolute pressure module.
ZERO_MEAS?
Shows the zero offset or the reference value for absolute pressure modules.