Fluke 3130 User Manual page 48

Portable pressure calibrator
Hide thumbs Also See for 3130:
Table of Contents

Advertisement

3130
Users Manual
LOCAL
Returns the Product to local operation if it was in remote mode. The command also clears
LOCKOUT if the Product was in lockout mode.
LOCKOUT
Sending this command sets the lockout state, when the unit is in REMOTE or goes to
remote, it stops use of the keypad completely. The lockout state can only be cleared by
sending the LOCAL command.
LO_ERR
Sets the 0 % point for the percent error mode calculation in the current engineering units.
For example:
To set the 0 % point to 20 psi, send LO_ERR 20.
LO_ERR?
Shows the 0 % point for the percent error mode calculation.
For example:
If the 0 % point is set to 20 psi, LO_ERR? shows 2.000000E+01, PSI .
MOTOR_ON
Turns on the motor.
MOTOR_OFF
Turns off the motor.
MOTOR?
Shows the state of the motor.
For example:
If the motor was on, MOTOR? shows ON.
40