WARNING!
Do not load the seat
Do not load the seat or the AP elevator during this operation. Any load on
the seat or the AP elevator could cause permanent damage to the
wheelchair or injuries on person(s) in the wheelchair or in its close vicinity.
These conditions apply until the screws are reinstalled and tightened to the
correct torque.
6. Remove the four screws holding the battery box.
7. Disconnect the right motor and inhibit cable connector (C).
8. Disconnect the left motor and the bus cable connector (B).
9. Disconnect the control panel connector (A).
Maintenance and repairs
Figure 182. The battery box is secured with four bolts.
Figure 183. Connectors (B) and (C); for disconnection.