Advanced Task
Flow Function
•
To apply the flow function to a selected input, enable FLOW from the process
menu and press ENTER. The square root symbol is displayed beside the input
value to indicate that the FLOW function is active:
% Span
•
To convert a selected input display from a numerical value to a percentage of full-
scale reading, enable SPAN from the process menu and press F1 (SETTINGS) to
provide access to the span definition parameters. Two span definitions are
required, Zero and Full Scale.
To leave span at current setting, press EXIT.
To define zero and full-scale settings, proceed as follows.
To cancel % SPAN, press INPUT and turn function OFF at the process menu.
To cancel FLOW, press INPUT and turn function OFF at
the process menu.
23
Process Functions
K0430 Issue No. 2