ERROR CODE
(None)
DEACTIVE displays and the
wheelchair will not drive.
(None)
MAX BACK ANGLE displays.
(None)
CONTROLLER INHIBITED
displays and the wheelchair does not
drive.
(None)
SLOW DOWN is displayed and
the wheelchair drives slowly.
Part No 1141471
SYMPTOM
If the wheelchair has a TIAM and
RIAM, this message is displayed
when the Tilt angle, the recline
angle or the combined tilt and
recline angle are greater than 20°
from fully upright (beyond the
drive lock-out angle of 20°).
If the wheelchair has a TRAM or
TIAM, this message is displayed
when the tilt switch is open.
The wheelchair back has reached
the maximum programmed back
angle on a wheelchair with a
TIAM, RIAM or TRAM.
The system is tilted or reclined
beyond the drive lock-out angle of
20°.
Incorrect ACC DCI setting in the
calibration menu.
The elevating seat is elevated.
PROBABLE CAUSE
89
User Settings/Diagnostics
SOLUTIONS
Return the system to upright position.
The wheelchair back will not go past the
programmed maximum back angle. This is
normal behavior.
Return the system to the upright position.
Return elevating seat to the lowest position to
drive at full speed.