Invacare MK6i Reference Manual page 74

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

Calibration Menu
CALIBRATION
TILT CALIBRATE
RECLINE CALIBRATE
CM LEGS CALIBRATE
NOTE: Not available on nXc joystick.
BACK ANGLE
NOTE: Not available on nXc joystick.
START IN DRIVE
NOTE: Not available on nXc joystick.
ATT PWR OVERRIDE
NOTE: Not available on nXc joystick.
PACM6 ADJUST
NOTE: Not available on nXc joystick.
DIGITAL ATTENDANT
CONTROL
NOTE: Not available on nXc joystick.
Invacare® MK6i™Electronics
LCD DISPLAY
TILT CALIBRATE
MOVE DOWN
SET DOWN ANGLE
MOVE UP
SET UP ANGLE
BACK ANGLE
95
START IN DRIVE
>>LAST USED
DRIVE 1
DRIVE 2
DRIVE 3
DRIVE 4
ATT PWR OVERRIDE
PACM6 ADJUST
DIGITAL ATTENDANT
CONTROL
Calibrating Tilt, Recline or Center Mount Legs, requires a Pitch-Angle
Gauge.
>
__°
• For Custom Actuators Only.
>
1.
Select MOVE DOWN.
__°
2.
Use the down arrow key to tilt the system all the way.
3.
Press the Select key.
4.
Select SET DOWN ANGLE.
5.
Measure the angle of the seat using the Pitch-Angle Gauge.
6.
Use the arrow keys to set the SET DOWN ANGLE to the gauge
measurement.
7.
Repeat for MOVE UP and tilt the system all the way up.
The angle of the back relative to the seat.
• The back angle plus the tilt angle determines the drive lockout angle.
• A value between 85° and 105°, typically set at 95°.
• Only displayed with smart actuators on tilt only systems.
Allows Setting the Drive Mode (1 through 4) the wheelchair powers up
into.
• RETURN TO LAST USED allows the wheelchair to power up into the drive it
was in when last powered down.
• DRIVE 1 allows the wheelchair to ALWAYS return to 1 when turned on.
Allows Setting the Attendant Power Overide Mode.
• ON, the wheelchiar can not be turned off unless the attendant control is also
off.
• OFF, the wheelchiar can be turned off regardless of the status of the attendant
control.
Provides access for programming all driving Performance Adjustment
Settings for the Proportional Attendant Control.
Provides access for programming all driving Performance Adjustment
Settings for the Digital Attendant Control.
74
DESCRIPTION
Part No 1141471

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents