• Each of the 4 drives can be programmed to use the G-Trac feature, or have G-Trac disabled in that drive. To
accomplish, use a Hand Held Programmer and select the Performance Adjustments menu item. Scroll down to G-
TRAC and set it for ON or OFF.
A function of Time & Power. How quickly programmed Power Level is reached.
• ALL FOUR quadrants are affected by the programmed torque level.
MORE
• High = immediate ramping up to programmed power level. Recommended for Slow Speeds, Switch drivers, rough
terrains, curbs, significant obstacles.
• Low = slower ramping up to programmed power level. Recommended for faster speeds, level terrains, new drivers
- even at some slower speeds.
• MK6i Torque levels are noted in Ohms.
• Small changes to torque have a significant impact.
• Recommendation is to change only in 4 Ohm increments to determine if needs are met.
Determines Drive Lock Out Function for systems utilizing any version of the SPJ+ Joystick
OFF
• Disables Drive Lock Out for tilt or recline systems (manual or power),
• Recommended setting for chairs without tilt or recline seating systems
NORMALLY OPEN/AFTERMARKET
• Setting for some systems using aftermarket powered seating systems. Consult the aftermarket seating manufacturer
for more information.
NORMALY CLOSED/INVACARE
• Setting used for Invacare tilt or recline systems (manual or power).
PERFORMANCE ADJUSTMENT DESCRIPTION
31
MK6i Performance Adjustments