Invacare MK6i Reference Manual page 32

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

MK6i Performance Adjustments
LCD DISPLAY
DCI MONITORING
(SPJ+ ONLY)
>IVC MANUAL SEATING
(OR "CONTINUOUS")
IVC POWER SEATING
(OR "LATCHING")
ACC 1 (2)
(SPJ+ ONLY)
OFF
ON
TRACTION
30%
LESS
|||||
JSTK THROW
MOVE JOYSTICK TO:
FORWARD_ REVERSE_
LEFT_ RIGHT_ AND
THEN NEUTRAL_
AXES SEL
FORWARD
>FORWARD
REVERSE
>REVERSE
LEFT
>LEFT
RIGHT
>RIGHT
Invacare® MK6i™Electronics
Determines the method Drive Lock Out status is monitored by the controller
CONTINUOUS/IVC MANUAL SEATING
• Status of the drive lockout switch is continuously monitored for change.
• Use this setting for manual tilt or recline systems.
• Recommended for some aftermarket powered seating systems. Consult the aftermarket seating manufacturer for
more information.
LATCHING/IVC POWER SEATING
• Monitors the Drive Lockout Input switch ONLY during power up, and after actuator operation
• This setting is recommended for Invacare powered seating systems
Allows setting actuator to operate through the SPJ+ with ACC Joysticks when ACC is used on the TDX
Spree Power Wheelchair with power tilt.
ACC 1 = ELEVATE, ACC2 = TILT
A reduction of the speed when going into and coming out of turns.
• The higher the value, the greater the reduction in speed.
MORE
• Set at 0% for the majority of wheelchair users.
• Increasing values may be helpful to soften veer correction in "latched" driving mode, or to dampen veer correction
speed for aggressive drivers.
Used to calibrate joystick throw.
• Sets the point for reaching full speed in relation to joystick displacement.
• Values DO NOT return to default settings unless Manually Re-Set.
• Used with individuals having reduced ROM available for joystick operation.
Assigns / Re-Assigns joystick commands to a desired direction.
• Useful when changing "Joystick Operation" in relation to "Joystick Position".
• Each of the four input axes can be redirected to any output axis, or turned off.
• Settings DO NOT refer back to default unless manually re-set.
• "Select" Key on Programmer cycles through output choices.
PERFORMANCE ADJUSTMENT DESCRIPTION
32
Part No 1141471

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents