Invacare MK6i Reference Manual page 66

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

Calibration Menu
CALIBRATION
ACC DCI
MONO PORT 1
MONO PORT 2
DISPLAY ORIENT
NOTE: Not available on nXc joystick.
Invacare® MK6i™Electronics
LCD DISPLAY
>>OFF
INVACARE MANUAL
(OR CONTINUOUS)
INVACARE POWER
(OR LATCHING)
MONO PORT 1
OFF
>>DRIVE SELECT
MODE SWITCH
<ACTUATOR> UP/DOWN
<ACTUATOR> UP
<ACTUATOR> DOWN
MONO PORT 2
OFF
>>DRIVE SELECT
MODE SWITCH
<ACTUATOR> UP/DOWN
<ACTUATOR> UP
<ACTUATOR> DOWN
DISPLAY ORIENT
NORMAL
INVERTED
Determines Tilt Switch Function for the Actuator assigned to the ACC
Controller (conventional Actuators and ACC Controllers only)
Allows turning Drive Lock out OFF on Single Actuator Systems or set-
ting how the Controller monitors the Drive Lockout Switch (only
when Actuator is operating through the ACC of the Controller)
Standard settings are:
• OFF: Standard for IVC Tilt and Recline combination systems, any system w/
Power Center Mount Leg or Tilt w/Elevate and systems with smart actuators.
Also allows disabling drive lockout for Conventional Single Actuator Systems
• IVC Manual (Continuous): Drive Lockout switch status is continually
monitored. Used with IVC Manual Tilt or Recline systems and some
aftermarket Powered Seating Systems
• IVC Power (Active): Formula CG Single Actuator powered seating systems
Assigns the Function of the Left Mono Port on the MK6i Display &
Multiple Drive Joysticks.
• Drive Select allows the mono switch to change Drives 1
through 4.
• Mode/Reset allows the Switch to function as a reset switch.
• <Actuator> Up/Down allows the switch to operate the actuator (up/down
mode) when one actuator is connected to the system.
• Mono Port 1 is the default if not using a "Y" cable (splitter).
Allows a second switch function in the Right Mono Port of the MK6i
Display & Multiple Drive Joysticks.
• If a second function is selected, a "Y" Splitter cable –Stereo to two mono – is
required to access the second switch port or else a stereo switch (2 PB, 2 WT)
may be used.
Only available when CMPJ+, PSF or PSR is on the wheelchair.
• Choose Normal for CMPJ+ or PSF.
• Choose Inverted for PSR.
66
DESCRIPTION
Part No 1141471

Hide quick links:

Advertisement

Table of Contents
loading