Invacare MK6i Reference Manual page 37

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

LCD DISPLAY
MOUSE AXES
FORWARD > VERT
REVERSE > R-CLK
LEFT > L-CLK
RIGHT > HORZ
MOUSE B
IR
> OFF
3 QUADRANT
4 QUADRANT
NO DRIVING
> ON
OFF
LIGHT CONTROL
> ON
OFF
Part No 1141471
PERFORMANCE ADJUSTMENT DESCRIPTION
Assigns / Re-Assigns driver control commands to desired mouse function.
Available ONLY when 3 Quadrant Mode is selected.
• Assignments are separate from driving axes selections.
• To allow "Right Click" as a choice for a mouse axis:
• Must have a 4 Quadrant Control.
• Choose "3 QUADRANT" in Mouse mode set-up.
• Reverse can then be set to "Right Click" if needed.
• VERT = Mouse movement toggles between Up and Down.
• R-CLK = Mouse Right Click.
• HORZ = Mouse movement toggles between Left and Right.
• L-CLK = Mouse Left Click.
For MOUSE "B" adjustments refer to MOUSE, MOUSE SETTINGS and MOUSE AXES.
• USB mouse associated with the RF Mouse Only Module for software versions 2.2 or higher.
Activates Control of Infrared Devices Enabled from Calibrations menu.
• OFF Disables IR Control for that drive.
• 3 Quadrant: Forward command disabled (Head Array Users can use back pad as head rest).
Right = scrolls through Menu.
Left = Selects Icon.
• 4 Quadrant: Left – Right scroll through icons, Forward Command Selects Icon.
Allows Driving to be turned off for that particular drive.
• Eliminates driving to dedicate that drive to another activity.
• Helpful when performing multiple activities through the driver control to reduce choices users need to make.
• Can be used to eliminate access to drives (until user is ready to add functions / features).
Allows lights to be turned on/off through the drive control in that drive.
37
MK6i Performance Adjustments
Invacare® MK6i™Electronics