Invacare MK6i Reference Manual page 83

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

ERROR CODE
E09 (LEFT)
LEFT PARK BRAKE FAULT displays
and wheelchair does not drive.
E10 (RIGHT)
RIGHT PARK BRAKE FAULT
displays and wheelchair does not
drive.
E14
Battery Fault displays and the
wheelchair does not drive.
E18
NEUTRAL TESTING displays.
E19
BAD JOYSTICK CAL VALUES
displays and the wheelchair does not
drive.
E28
CHARGER PLUGGED IN displays.
E32
JOYSTICK TIMEOUT displays and
the wheelchair does not drive
E41
CONTROLLER STARTUP FAULT
displays and the wheelchair drives
slowly.
E102
GB GRNL FAULT
Part No 1141471
SYMPTOM
Motor, Controller or other
electrical device (Error code E9 or
E10).
Motor, Controller or other
electrical device (Error code E9 or
E10).
The controller has determined the
batteries need to be replaced.
The joystick neutral test has failed.
The joystick calibration values are
outside of the expected range.
Battery charger connected.
Joystick or input device is
disconnected.
The controller has determined a
fault during a previous turn-off
process.
Unidentifiable Error
PROBABLE CAUSE
83
User Settings/Diagnostics
SOLUTIONS
Ensure motor lock/levers are engaged before
turning power on.
Call Technical Services.
Replace batteries.
Release the joystick and try to get the joystick
back into the center-most position.
Recalibrate the joystick (joystick throw
procedure).
Unplug battery charger from the wheelchair if
charging is complete.
Turn off power, reconnect the joystick of input
device and turn power on.
Turn the wheelchair off and back on.
Call Technical Services.
Invacare® MK6i™Electronics