Invacare MK6i Reference Manual page 85

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

ERROR CODE
E140, E141
CTLR PWR FAULT
E142
LOW BATTERY
E143
HI BATT VOLTS
E144
M2 MTR FAULT
E145
M1 MTR FAULT
E146-E150
GB CTRL FAULT
E151, E152
M2 MTR FAULT
E153, E154
M1 MTR FAULT
E155, E157
GB CTRL FAULT
Part No 1141471
SYMPTOM
Check Joystick Cabling
Low battery
High Battery Fault
Left Motor (on RWD)/Right
Motor (on CWD) brake coil
short circuit
Right Motor (on RWD)/Left
Motor (on CWD) brake coil
short circuit
GB Controller Failure
Left Motor (on RWD)/Right
Motor (on CWD) Hall Sensor
Fault
Right Motor (on RWD)/Left
Motor (on CWD) Hall Sensor
Fault
Current Calibration Lost
PROBABLE CAUSE
85
User Settings/Diagnostics
SOLUTIONS
Check all connections for physical damage.
Check Joystick and Joystick Cabling.
Recharge batteries.
Replace batteries if not corrected after
charging.
Check Battery Voltage.
Call Technical Service.
Check Left Motor (M2) (on RWD)/Right
Motor (on CWD) Cabling.
Check Left Motor (M2) (on RWD)/Right
Motor (on CWD) Cabling.
Replace controller and recalibrate motors
Call Technical Service
Check Left Motor (M2) (on RWD)/Right
Motor (on CWD) Cabling.
Check Right Motor (M1) (on RWD)/Left
Motor (on CWD) Cabling.
Turn chair off and then back on.
If error repeats, replace controller and
recalibrate motors.