Invacare MK6i Reference Manual page 88

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

User Settings/Diagnostics
ERROR CODE
E203
RIGHT MOTOR FAULT
displays and the wheelchair does not
drive.
E204
REMOTE FAULT displays and
the wheelchair does not drive.
E205
CONTROLLER FAULT
displays and the chair does not drive.
E206
CONTROLLER WRONG
REMOTE displays and the wheelchair
does not drive.
E207
CONTROLLER SETUP FAULT
displays and the wheelchair does not
drive.
E208
G-TRAC FAULT is displayed
Invacare® MK6i™Electronics
SYMPTOM
Displayed when a problem with
the right motor is detected.
Right on RWD - Left on CWD
This is displayed when the
controller determines an
incorrect configuration.
This is displayed when the
controller fails a power-up test.
This is displayed when he
controller has determined an
invalid configuration.
This is displayed when the
controller module does not
recognize the MK6 CMPJ+/Display
as a valid device.
The G-Trac module or controller
is not functioning correctly.
PROBABLE CAUSE
88
SOLUTIONS
Check motor lock engagement (clutch).
Check motor connection plug.
Verify left/right by switching motor plugs.
If fault follows motor, replace motor.
If fault does not follow motor, replace
controller.
Replace the controller.
Replace the controller.
Check all connections and wiring.
Replace the controller.
Replace the G-trac module and/or the
controller. Please note that the chair will drive
with this error displayed, however, the G-trac
feature is disabled and the chair performs
without the benefit of the G-Trac features.
Part No 1141471