Invacare MK6i Reference Manual page 84

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

User Settings/Diagnostics
ERROR CODE
E103
GB FAULT - CYCLE PWR
E104-E105
GB CTRL FAULT
E106-E107
GB CTRL FAULT
E108-E109
CURR CAL FAULT
E110-E111
GB CTRL FAULT
E112-E113
GB CTRL FAULT
E114-E127
CURR CAL FAULT
E128
M2 MOTOR FAULT
E129
M1 MOTOR FAULT
E130, E132
M2 MTR CAL
E131, E133
M1 MTR CAL
E134-E139
SW FAULT
Invacare® MK6i™Electronics
SYMPTOM
Possible Controller Failure
Left Current Sensor Error
Right Current Sensor Error
Current Calibration Error
Left (on CWD)/Right (on RWD)
Current Sensor Error
Right (on RWD)/Left (on CWD)
Current Sensor error
Current Calibration Error
Left Motor (M2) (on RWD)/Right
(on CWD) Over Current Fault
Right Motor (M1) (on RWD)/Left
(on CWD) Over Current Fault
Left Motor (RWD)/Right Motor
(CWD) - Too much drag/load
Right Motor (RWD)/Left Motor
(CWD) - Too much drag/load
Controller Software Fault
PROBABLE CAUSE
84
SOLUTIONS
Turn chair off and then back on.
If fault repeats, replace controller and
recalibrate motors.
Replace controller and recalibrate motors.
Replace controller and recalibrate motors.
No action required - Factory Test Only
Call Technical Services.
Replace controller and recalibrate motors.
Call Technical Services
Replace controller and recalibrate motors.
No action required - Factory Test Only
Check Left Motor (M2) (on RWD)/Right
(on CWD) and Cabling.
Check Right Motor (M1) (on RWD)/Left
(on CWD) and Cabling.
Recalibrate motor.
Recalibrate motor.
Replace the controller.
Call Technical Service.
Part No 1141471

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents