Invacare MK6i Reference Manual page 72

Remote programmer
Hide thumbs Also See for MK6i:
Table of Contents

Advertisement

Calibration Menu
CALIBRATION
4 WAY SWITCH
HARD PUFF
SOFT PUFF
HARD SIP
SOFT SIP
NOTE: Not available on nXc joystick.
Invacare® MK6i™Electronics
LCD DISPLAY
4-WAY SWITCH
FWD
-TILT U/D
REV
-RECL & LEGS U/D
LT
-ELEVATE UP
RT
-ELEVATE DOWN
HARD PUFF CAL.
MIN
.70 MAX
S
||
Allows customization of the operation of the 4 way switch to meet
users needs.
• Select switch quadrant (FWD, etc.) to view list of choices.
• Each actuator function listed below has a choice for U/D (Up/Down), Up or
Down and corresponding icons.
OFF
TILT
RECL
LEG
ELEVATE
RIGHT LEG
LEFT LEG
RECL & LEG
Calibrates pressures required to activate hard / soft, puff & sip of com-
mands.
H
• Separate screens for each of the 4 pressures.
• Select command to calibrate. Follow instructions below. Save on completion.
• GOAL: Separate "S" & "H" values sufficiently for easy distinction between
Hard & Soft commands – AND set values low enough to assure they can be
consistently achieved. Values should be between .10 and 1.28.
• In a Hard calibration mode, use the up/down arrows to raise and lower the H
value which MUST be met as the user puffs/sips.
• In a Soft calibration mode, use the up/down arrow keys to raise and lower the
S value, which MUST be met as the user puffs/sips.
72
DESCRIPTION
Part No 1141471