Section 5.5 Interface Test - Welch Allyn Spot Vital Signs Service Manual

Hide thumbs Also See for Spot Vital Signs:
Table of Contents

Advertisement

Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
1. Connect the unit to the test station by hooking up the pneumatic tubing, removing the battery and
connecting the power supply. Ensure that there are no obstacles in the way of the IR window.
2. Select Test|Display in the repair software.
3. Select "all on" in the LCD Display section and ensure that all the segments are lit on the unit.
4. Select "all off" in the LCD Display section and ensure that all the segments are not lit on the unit.
5. Select "normal" in the LCD Display section to return the display to normal.
6. Select "on" in the Back-light section and ensure light comes on.
7. Select "off" in the Back-light section and ensure light goes off.
8. Select the buzzer test and ensure that the buzzer of the Spot Device is "on." Select "OK" to exit
test and turn off the buzzer.
9. Select "OK" to exit dialog box.

Section 5.5 Interface Test

End of Section 5.5
Page 51 of 64