Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
Press the "start" button to close the valve. Pump up the unit using the bulb and valve
inflate the unit to 250 mmHg. Confirm the unit meets the specification for leak called out
in the Repair Test Specification document. If the unit is leaking:
E. Pinch off the tubing leading to the pump. If this stops the leak, change the pump.
(Reference section 4.8 page 46)
F. Pinch off the tubing leading to the valve. If this stops the leak, change the valve.
(Reference section 4.8 page 46)
G. Pinch off the tubing leading to the transducer. If this stops the leak, change the Main PCB.
(Reference section 4.6 page 44)
7. If the unit is inflating/deflating properly however there is no reading given, change out the main
PCB. (Reference section 4.6 page 44)
8. FOR THE SPO2 OPTION: Connect the patient simulator to the unit SpO2 connection by the
Nellcor SpO2 cable. Confirm readings equal to the specifications called out in the Repair Test
Specifications document. If there is no reading, change out the SpO2 PCB. (Reference section 4.7
page 45)
9. FOR THE TEMPERATURE OPTION: Remove the temperature probe and view the
temperature window on the display. Make sure the temperature is in the "monitor mode." This will be
displayed at the bottom of the window. If the unit is not in monitor mode, wait approximately 40-60
seconds and the unit will automatically transfer into the monitor mode. Using the Diatek 9600
calibrator set at 96.4 F/ 35.8 C or using a water bath between 84 F and 106 F, place the probe into
the small hole on top of the 9600 calibrator or into bath. If there is no reading place the probe back
into its housing, remove and retry. If there is still no reading, change the probe and retry the test. If
there is still no reading place the probe back into the housing, unplug the probe and plug in the cal-
key (5200-25), remove the probe and Confirm reading meets specifications called out on the Repair
Test Specification document. If there is no reading at this point, change the temperature connector
PCB and cable to the main PCB. (Reference section 4.2 page 40)
Page 27 of 64