TEMPERATURE BOARD ERROR CODES
ERROR CODE
E0.1
E0.2
E0.3
Show "A" in Temperature display. Flash Up arrow.
Show "A" in Temperature display. Flash Down arrow.
E0.6
E0.9
E1.0
E1.1
E1.2
E1.4
E1.5
E1.6
E1.7
E1.8
E9.1
Broken probe icon:
"P"
C22
DESCRIPTION
Probe Heater Accumulator
Probe A/D Pulse Width Error
Adaptive Probe Gain too High or Low
Ambient Temperature too High
Ambient Temperature too Low
Host system has violated the interface rules
Rcal resistor pulse too long.
Rcal resistor pulse too short
PTB Resistor above 98.55ºF
PTB Resistor below 98.15ºF
Q Only turned heater circuit on
C only turned heater circuit on
Heater Circuit Error
Heater temperature error
Heater found on at end of pulse.
Communication Error
Broken or Missing Probe
Loss of Tissue Contact
10 Minute Diagnostic Limit Exceeded
End of Section 3.4