Welch Allyn Spot Vital Signs Service Manual page 68

Hide thumbs Also See for Spot Vital Signs:
Table of Contents

Advertisement

Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
General Notes
Note 1: The word "Unit" throughout this document refers to the Spot Vital Signs with out
pneumatics (tubing and cuff), temperature probe, SpO2 probe, and main battery attached, unless
other wise noted.
Note 2: Standard test voltage unless otherwise stated is 6.5 (+0/- 0.25) Vdc.
Note 3: All test are performed utilizing the Repair Software for the Spot Vital Signs or following the
procedures called out in the Spot Service Manual.
Note 5: Unless otherwise stated, all calibrated volumes will be +/- 10cc of the stated volume.
Note 6: Unit software for the Spot Vital Signs will be equal to or greater than "1.00" for "Spot"
software (SW 1) and equal to or greater than "1.00" for ModB software (SW2). If the unit has
Nellcor SpO2, the SpO2 software will be equal to or greater than "1.2.0.0." If the unit has
temperature, the temperature software will be equal to or greater than "2.1."
Page 61 of 64