Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
1. Connect the unit to the test station by hooking up the pneumatic tubing, removing the battery and
connecting the power supply. Ensure that there are no obstacles in the way of the IR window.
2. Choose Test|Button Test. Press each button and ensure that the computer acknowledges it, via
the "Button Pressed" display.
3. Select "OK" to exit from dialog box.

Section 5.4 Button Test

END OF SECTION 5.4