Section 4 Removal/Replacement Of Parts; Battery - Welch Allyn Spot Vital Signs Service Manual

Hide thumbs Also See for Spot Vital Signs:
Table of Contents

Advertisement

Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A

Section 4 Removal/Replacement of Parts

4.1 Battery

1. Lay unit face down on the test bench
2. Remove the 4 Phillips head screws holding the battery cover/handle to the unit using T112918
with the #4 Phillips bit.
3. Lift handle/cover and aside.
4. Lift battery from housing and disconnect the Molex connector.
Reassemble in reverse order.
End of Section 4.1
Page 39 of 64