Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
Test Information
Unit SW Version
SpO2 SW Version
Test Pressure @ 275 mmHg
Unit pressure @ 275 mmHg
Test Pressure @ 175 mmHg
Unit Pressure @ 175 mmHg
Test Pressure @ 50 mmHg
Unit Pressure @ 50 mmHg
Test Pressure @ 0 mmHg
Test Pressure @ 0 mmHg
Test Voltage
Unit Voltage
Blank Current
Back Light Current
Valve/Pump Current
SpO2 Current
Noise Level
Button Test
Interface Test
Print Quality (If applicable)
Unit Leak
Unit Dump Time
Unit Inflation Time
Valve Control Test
Over Pressure Test
Over 15 mmHg Test
SpO2 Simulator Reading
SpO2 sensor Reading
Temperature Calibration Reading (96.4 F))
Temperature Calibration Reading (106 F)
Temperature Cal Key Reading
Test Data
Test Specifications
+/- 5 mmHg
+/- 3 mmHg
+/- 5 mmHg
+/- 3 mmHg
+/- 5 mmHg
+/- 3 mmHg
+/- 1 mmHg
+/- 1 mmHg
6.0 + 0/-0.2 Vdc
+/- 0.05 Vdc
<= 200 mA
<= 400 mA
<= 700 mA
<= 120 mA
<= 0.05 mmHg
Pass/Fail
Pass/Fail
Pass/Fail
<= 5 mmHg in 15 seconds.
<= 10 seconds
<= 7 seconds
Pass/Fail
296 mmHg to 329 mmHg
<= 180 seconds
81% +/- 3%; 112 BPM +/- 3 BPM
Pass/Fail
96.4F+/-0.3F (35.8C+/-0.2C)
106F+/-0.3F (41C+/-0.2C)
97.3F+/-0.2F (36.3C+/-0.1C)
Page 58 of 64