Welch Allyn Spot Vital Signs Service Manual page 14

Hide thumbs Also See for Spot Vital Signs:
Table of Contents

Advertisement

Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
Overpressure Cutoff
315 mmHg ±15 mmHg
Temperature Ranges
Temperature Accuracy
Temperature accuracy meets or exceeds ASTM E112-86: "Standard Specification for Electronic
Thermometer for Intermittent Determination of Patient Temperature."
Temperature Determination Time
(Oral) 4 seconds typical, 15 seconds maximum
(Axillary) 10 seconds typical
(Rectal) 15 seconds typical
Oxygen Saturation Range (SpO2 %)
40-100% oxygen saturation
SpO2 Accuracy
±3% in the range of 70-100% oxygen saturation (1 Standard Deviation)
<70% unspecified by the OEM.
Battery Charging
To 90%-100% capacity in 12 hours. Unit will operate and charge battery simultaneously when
connected to power source.
Mechanical Specifications
Dimensions
Height 9.70 inches (24.64 cm)
Length 5.72 inches (14.53 cm)
Depth 4.73 inches (12.01 cm)
Page 14 of 64