Self Test And Service Menus - Welch Allyn Propaq Encore Service Manual

Vital signs monitor
Hide thumbs Also See for Propaq Encore:
Table of Contents

Advertisement

Service Manual

Self Test and Service Menus

The Encore monitor includes built-in self-tests that verify the integrity of its hardware and
software each time you turn it on. If the monitor turns on and displays the Main Menu, all
self-tests were passed. If a self-test results in a failure, a message and error code are
displayed indicating the failure mode. Refer to
Welch Allyn Technical Support for help.
The Encore monitor contains software routines that make the Functional Verification and
Calibration procedures quicker and easier. You access these routines through the Service
Menus as indicated in each of the procedures.
The figure below shows the hierarchy of the Service Menus once you enter the top level
Service Menu.
The procedures in this manual indicate which buttons to push by separating each button
menu name with the (>) character. For example, to access the Service Menu, the
procedures would indicate for you to press SETUP > MORE > MORE > SERVICE > YES,
which means you press the SETUP button, then the MORE button, then the MORE
button, then the SERVICE button, and finally the YES button.
Note
service menus
(Software v. 2.5X only)
The figure illustrates all Service Menu functions. Some of these functions are not
available if you do not have the appropriate option.
NIBP TEST
IBP TEST
PUMP TEST
AUTO PMP
TEMP TEST
PIXL TEST
CYCLE
CHECKER 1
CHECKER 2
LOOP TEST
MAIN CAL
MAIN TEST
SIDE CAL
SETTINGS
KEY TEST
KEY 0
KEY 1
NEXT
CHANGE
"Troubleshooting"
SpO2 TST
CO2 TEST
NIBP CAL
ZERO
PREV MENU
NET TST
BORDER
PREV MENU
OUT TEST
PREV MENU
SIDE TEST
PREV MENU
SYSTEM
KEY 2
KEY 3
PREV MENU
ALL ALARM
PREV MENU
Overview
on page 43 and contact