Welch Allyn, Inc. Spot Vital Signs Service Manual
Revision A
10. FOR THE COMMUNICATION OPTION: Select Tools|Options to insure that the
communication port that the cable is connected to is selected. Select "test" and ensure that the Spot
Device responds accordingly. If an error is present, ensure that there is no obstacles in the way of the
IR window and re-try the test. If the error is still present, change out the main PCB. (Reference
section 4.6 page 44)
End of section 3.1
Page 28 of 64