Power on, power-on self-test, and power off

When the battery is charged, press
The monitor runs a diagnostic self-test each time it powers up.
8
8
When the self-test is complete, the software version appears briefly in the message
window, followed by the current time of day.
To shut off the monitor, press
Note
If all tested functions are working normally, the various windows briefly display
start-up values ('8' and '- -') and a short tone sounds twice.
Caution Users should check for audible alarm function every time the VSM 300
is used. During the normal power-up cycle, two audible tones are emitted
immediately after the self-test is complete. If these tones do not sound, the audio
has failed. Remove the device from service and contact Welch Allyn.
The loss of the audible alarm could cause a delay in a clinician learning of an alarm
condition for the following conditions: 1) hypotension or hypertension, 2) low
blood oxygen content (SpO
relating to the loss of monitoring of a patient (e.g., a "sensor off" condition). Such
delay could potentially result in injury to the patient.
mmHg
kPa
SYS
SpO
%
2
8
8
.
.
.
DIA
8
8
.
.
.
****************
SYS
SpO
%
2
1
.
DIA
****************
If the self-test fails, an error code appears in the SYS window.
Caution Always observe the monitor during power-up. If any display fails to
illuminate properly, or if an error code appears in the systolic window, inform your
biomedical engineering department immediately, or call your nearest Welch Allyn
Customer Service or Technical Support facility. Do not use the monitor until the
problem is corrected.
Shutting off the monitor erases all stored patient data but does not erase settings
or configuration parameters.
to turn on the monitor.
), 3) low or high pulse rate, 4) other alarm conditions
2
/min
SYS
DIA
1
2
3
4
/min
mmHg
8
8
8
SYS
.
.
.
ºF
ºF
8
8
8
ºC
ºC
.
.
.
DIA
M
.
Chapter 2 Setup
/min
SpO
%
2
18
8
.
.
****************
/min
SpO
%
2
0
0
00:00:45
15
0
ºF
ºF

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

530005300p530t0530tp53s0053st0 ... Show all

Table of Contents