Preparation of Existing Systems

Although upgrading Chemstation and the operating software of the mass
spectrometer will not delete any user data if performed correctly, Agilent
highly recommends that you backup any methods, sequences, user data and
mass spectrometer tune files before starting the software upgrade.
Make sure the customer has produced current and functional backups of all user data.
C A U T I O N
This includes raw data files, methods, sequences and MS tune files. Backing up user
data is not covered by the Chip Cube installation.
1 Optimize the Agilent 1260 Infinity stack configuration as described in
Allow 36 cm (14") space between Agilent 1260 Infinity stack and mass spectrometer for the
N O T E
Chip Cube.
2 Update the firmware of all Agilent 1260 Infinity modules to revision
Agilent 1260 Infinity firmware revision A.06.01 is incompatible with earlier revisions of
C A U T I O N
firmware. In systems with mixed revisions A.06.xx and A.05.xx, certain modules
become invisible and cannot be accessed.
See the Chip Cube Service Manual for details on how to recover from such situations.
HPLC-Chip Cube Interface User's Guide
"System Setup Considerations"
A.06.01 or higher. The firmware update program and the necessary
firmware files can be found on the Chemstation B.01.03 CD-ROM in the
\Support\fwupdate\ directory.
Preparation of Existing Systems
on page 42.