Approximately 15 cm (6 inches) of clear space behind the instrument is
required for air circulation and electric connections. One end of the power
cord must always be accessible.
For detaching the G4240A Chip Cube from the Chip Cube Interface on the
mass spectrometer, allow 10 cm (4 inches) of free space between the Chip
Cube and the LC stack. Agilent recommends stacking the HPLC modules in
either of the two configurations shown in
starting on page 34

Environment

Your G4240A Chip Cube will work within the specifications of ambient
temperatures and relative humidity described in
However, mass spectrometers commonly have tighter operating temperature
and humidity limits and therefore determine the applicable specifications.
Do not store, ship or use the G4240A Chip Cube under conditions where temperature
C A U T I O N
fluctuations could cause condensation within the G4240A Chip Cube. Condensation
will damage the electronics. If your G4240A Chip Cube was shipped in cold weather,
leave it in its box and allow it to warm up slowly to room temperature to avoid
condensation.
In order to function properly, do not place the G4240A Chip Cube on a vibrating surface
N O T E
or near vibrating objects.
The G4240A Chip Cube is designed to operate in a controlled electromagnetic
N O T E
environment (EN61326/A1) where RF transmitters, such as mobile phones, should not be
used in close proximity.
HPLC-Chip Cube Interface User's Guide
Site Requirements for the G4240A Chip Cube
Chapter
3, "Physical Specifications,"
Table 3
Site Preparation
on page 34.
33
3