Installation Procedures - Mindray BA-88A Service Manual

Semi-auto chemistry analyzer
Hide thumbs Also See for BA-88A:
Table of Contents

Advertisement

Databse: SqlServer 2005 Express
Network communication: 10/100M 10/100M
Communication interface: RS-232C
Hard disk ( the disk on which the software is installed): 10G
Printer (connected with PC)
2.2

Installation Procedures

1
When you receive the system, carefully inspect the package. After opening
the package, check the delivered goods against the packing list.
2
Move the instrument to the installation site and remove all the package and
protective materials.
3
Connect the instrument to waste container.
4
Connect the power cable and power on the analyzer. After self-check and
tubing wash is completed, the main screen is displayed.
5
After the system is stable, request one or two routine tests and run. Assess
the test results. Please refer to the operator's manual for details.
6
Follow the normal procedure to shutdown the analyzer.
7
Training:
Can
Is the customer familiar with the daily, weekly and monthly maintenance and
relevant maintenance methods?
Is the customer familiar with cleaning and washing the system?
Is the customer familiar with troubleshooting the common failures?
Is the customer familiar with replacing the peristaltic pump tubing and
calibrating the flow volume?
Is the customer familiar with replacing the aspiration tubing? Is the customer
familiar with replacing the lamp?
2-2
the
customer
2 System Installation
complete
daily
Yes □
Yes □
Yes □
Yes □
Yes □
Yes □
tests?
No □
No □
No □
No □
No □
No □