Flow Cell Base Assembly - Mindray BA-88A Service Manual

Semi-auto chemistry analyzer
Hide thumbs Also See for BA-88A:
Table of Contents

Advertisement

NOTE
Wear clean cotton gloves, when replacing the filter or lens. If contaminant is
found on the surface of the filter or lens, clean it with ethanol.
NOTE
The assembly or disassembly of the filter wheel, motor or home position
sensor should be followed with home position deviation adjustment.
Otherwise, the test result might not be reliable.
The home position deviation adjustment of the filter wheel should be followed
with the adjustment of photoelectric gain. Otherwise, the test result might not
be reliable.
3.2.4

Flow Cell Base Assembly

3.2.4.1 Introduction
The flow cell, temperature control elements, analogue board and etc are installed and
positioned by the flow cell base assembly. Two modes—flow cell and cuvette are
available on the system. Manual replacement is supported. The temperature control
system is composed of semi-conductor refrigeration plate, air duct, temperature
sensor and insulation material, providing 37℃, 30℃ and 25℃ which are needed on
the system. Analogue board collects the data of absorbance and temperature.
3.2.4.2 Components and Structure
The structure of the flow cell assembly is shown in the following figure.
3 System Description
3-9