Mindray BA-88A Service Manual page 14

Semi-auto chemistry analyzer
Hide thumbs Also See for BA-88A:
Table of Contents

Advertisement

Maintaining the System
Setting up the System
8
CAUTION
Maintain the system strictly as instructed by this manual. Inappropriate
maintenance may lead to unreliable results, or even equipment damage
and personal injury.
To wipe off dust from the system surface, use a soft, clean and wet (not too
wet) cloth, soaked with mild soap solution if necessary, to clean the
surface. Do not use such organic solvents as ethanol for cleaning. After
cleaning, wipe the surface with dry cloth.
Switch off all the powers and unplug the power cord before cleaning. Take
necessary measures to prevent water ingression into the system,
otherwise it may lead to equipment damage or personal injury.
Replacement of such major parts as lamp assembly must be followed by a
calibration.
Check the pump tubing for leakage as needed and replace the tubing in
time. Otherwise, the normal aspiration of the system might be affected. It is
recommended that the inner system tubing should be replaced every 24
months to avoid possible blockage or invalidation brought about by aging.
CAUTION
To define such parameters as calculation method and wavelength, follow
the instructions in this manual and the package insert of the reagents.
Foreward