Replacing Lamp - Mindray BA-88A Service Manual

Semi-auto chemistry analyzer
Hide thumbs Also See for BA-88A:
Table of Contents

Advertisement

2
Connect the thicker end of the new tube to the inlet metal tube and then guide
the tube through the deflection tube.

5.3.3 Replacing Lamp

When the system warns that the background is too weak (no bubble in the flow cell)
and the warning can not be dealt with by cleaning the flow cell, the optical element
and adjusting the photoelectric gain. The detailed operation is as follows:
BIOHAZARD
When replacing the pump tubing, wear gloves and lab coat and, if
necessary, goggles. In case your skin contacts the waste, follow
standard laboratory safety procedure and consult a doctor.
NOTE
After replacement is completed, please make sure that the flow cell
is pressed to the bottom of the installation hole (the top of the flow
cell should be on the same level with the installation surface).
Otherwise the test result might not be reliable.
NOTE
When connecting the tube, the tube should reach the root of the
metal tube to ensure reliability.
After the aspiration tubing is installed, the bending part should be
transitioned smoothly, no sharp angle, corrugation is allowed
5 Service and Maintenance
The
thicker
end
is
connected
with the metal
tube
Aspiration
deflection
tube
5-17