5.3.1 Replacing the tubing

The tubing is a component that should be replaced regularly. When the instrument is
in use, if you notice that the instrument can not aspirate or aspiration volume
decreases obviously, check for leakage in the peristaltic pump. If yes, the pump
tubing might be broken and it should be replaced. A pump tubing is in the accessory
package (outer diameter 3.2mm, length 150-160mm, yellow).The procedure for
replacing is shown as follows:
1
Shutdown the instrument and disconnect the power cable between the
instrument and the power supply.