1 Remove the cell holder form the multicell transport by unscrewing the cell
2 Turn the stirring module with the fittings in the desired direction.
3 Insert the connector of the stirring module either in the drilling under the light
Figure 5
Second mounting position
Installing and Operating Your Multicell Transport
Installing the Stirring Module
Installing the Stirring Module
The stirring module (part number G1120-60006) is an optional accessory for
the multicell transport. The stirring module comprises the stirring assembly,
four tubing clamps and 2 meters of 3/8 inch i.d. flexible tubing.
Two positions for mounting the stirring assembly are available. One under
the light path, as shown in Figure 5 and one under the cell at the left side of
the cell in the light path. You can turn the stirring assembly by 90 degrees
about the axis perpendicular to the base plate of the multicell transport, thus
bringing the fittings in three different positions.
Mounting the Stirring Assembly
holder screws.
path or in the drilling under the cell on the left hand side of the light path.
Mounting the Stirring Assembly
18
Inlet fitting (D)
Outlet fitting (E)
Stirring module mounted
under the light path
Cell holder screws