Diagnostics and Troubleshooting

Error Messages

Error Messages
There is only one error message that can occur using the multicell transport
with the Agilent ChemStation software.
Multicell Transport home position not found
The multicell transport mechanism has jammed.
Check the two screws that fix the multicell transport in the
spectrophotometer are not pushed up into the path of the mechanism (for
example, this is the case when putting the transport beside the instrument).
Make sure the carriage can move freely along its entire path, refer to
"General Troubleshooting Hints" on page 29.
The electronics have failed.
Check for defective HOME-switch, defective multicell transport cable or
defective motor, see "Checking the Electronics" on page 31.