Agilent Technologies 8453 Operator's Manual page 16

Uv-visible spectrophotometer
Table of Contents

Advertisement

Figure 3
Multicell transport
connector
4 Turn on the spectrophotometer at the lower-left part of the instrument, if not
5 Start your Agilent ChemStation software, if not already running.
6 To verify proper installation of the transport, use your operating software to
1 Cut the tubing into two pieces of desired length.
2 Connect the tubing to the inlet and outlet fittings. Connect the outlet of the
Installing and Operating Your Multicell Transport
Installing Your Multicell Transport
Connector at Rear of Multicell Transport
already running. The transport initializes, that is, the transport moves to the
HOME-position.
move it to every possible cell position.
Connecting the Water Supply Tubing for the Cell Holder
The transport is shipped with 4 tubing clamps and 2 meters of 3/8 inch i.d.
flexible tubing.
thermostatted water bath to inlet fitting (A) on the cell holder and connect the
outlet fitting (B) on the cell holder to the inlet of the thermostatted water bath.
16