Table 1
Description and Specifications
Specifications
Specifications
The multicell transport generally operates under the same environmental
conditions as those specified in the Reference Handbook for the Agilent 8453
spectrophotometer. A maximum water pressure of approximately 3.5 bar
(50 psi) can be supplied to the thermostatted cell holder and to the
water-driven stirrer.
Multicell Transport Performance Specifications
Item
Time to move between adjacent cell
positions
Position repeatability
Step resolution
Temperature range for thermostatting
(cell holder only)
Temperature difference between any
two cell positions
10
Specification
Comment
< 1 second
± 0.1 mm
Linear position error in moving to
a cell position within 1 hour
0.060 mm
5–90 ° C
0.3 ° C
With a water bath at 37 ° C